Journal of Radioanalytical and Nuclear Chemistry, ( ISI ), Volume (295), No (1), Year (2013-1) , Pages (157-162)

Title : ( Shielding optimization in an in vivo activation analysis setup )

Authors: yasaman rezaei moghaddam , Laleh Rafat Motavali , Seyyed Hashem Miri Hakimabad ,

Access to full-text not allowed by authors

Citation: BibTeX | EndNote

Abstract

A neutron irradiation setup for in vivo activation analysis was characterized to optimize its spectrum measurements. In order to have a reliable spectrum, some shields and filters were defined to eliminate as much background possible. Since there were different types of background sources, different states were studied and necessary shields were employed to optimize the spectrum. To use a neutron shield, boron carbide layers (3 cm) around 3 9 3 in. sodium iodine detectors were employed. To prevent c-ray background, bismuth (2 cm) is used under the detectors.

Keywords

, IVNAA  Background  Neutron shield  c, Ray filter
برای دانلود از شناسه و رمز عبور پرتال پویا استفاده کنید.

@article{paperid:1036791,
author = {Rezaei Moghaddam, Yasaman and Rafat Motavali, Laleh and Miri Hakimabad, Seyyed Hashem},
title = {Shielding optimization in an in vivo activation analysis setup},
journal = {Journal of Radioanalytical and Nuclear Chemistry},
year = {2013},
volume = {295},
number = {1},
month = {January},
issn = {0236-5731},
pages = {157--162},
numpages = {5},
keywords = {IVNAA  Background  Neutron shield  c-Ray filter},
}

[Download]

%0 Journal Article
%T Shielding optimization in an in vivo activation analysis setup
%A Rezaei Moghaddam, Yasaman
%A Rafat Motavali, Laleh
%A Miri Hakimabad, Seyyed Hashem
%J Journal of Radioanalytical and Nuclear Chemistry
%@ 0236-5731
%D 2013

[Download]