Title : ( A comparative study on neutron-gamma discrimination with NE213 and UGLLT scintillators using zero-crossing method )
Authors: E. Bayat , N. Divani-Vais , M. M. Firoozabadi , Nima Ghal-Eh ,Access to full-text not allowed by authors
Abstract
The neutron-gamma discrimination capability of the new-type UGLLT liquid scintillator has been compared with NE213 (or its Bicron equivalent, BC501A). An Am-Be combined neutron-gamma isotopic source and a zero-crossing circuitry were used to explore various biases resulting the optimum figure of merits (FOMs) of 1.52 and 1.18 for NE213 and UGLLT at 170 keVee bias, respectively. The study shows that the UGLLT scintillator can be a good alternative for NE213 in pulse-shape discrimination experiments.
Keywords
, Scintillator Neutron, gamma discrimination UGLLT PSD@article{paperid:1063611,
author = {E. Bayat and N. Divani-Vais and M. M. Firoozabadi and Ghal-Eh, Nima},
title = {A comparative study on neutron-gamma discrimination with NE213 and UGLLT scintillators using zero-crossing method},
journal = {Radiation Physics and Chemistry},
year = {2012},
volume = {81},
number = {3},
month = {March},
issn = {0969-806X},
pages = {217--220},
numpages = {3},
keywords = {Scintillator
Neutron-gamma discrimination
UGLLT
PSD},
}
%0 Journal Article
%T A comparative study on neutron-gamma discrimination with NE213 and UGLLT scintillators using zero-crossing method
%A E. Bayat
%A N. Divani-Vais
%A M. M. Firoozabadi
%A Ghal-Eh, Nima
%J Radiation Physics and Chemistry
%@ 0969-806X
%D 2012