Journal of Cancer Research and Therapeutics, Volume (15), No (1), Year (2018-1)

Title : ( Assessment of radiation leakage from treatment applicator of Siemens Primus Plus and Siemens Artiste linear accelerators )

Authors: seyede motahareh zohur karbaf barbari , Mohammad Hadi Hadizadeh Yazdi , Mahdi Ghorbani , Sara Abdollahi ,

Citation: BibTeX | EndNote

Abstract

ABSTRACT Aim: The purpose of this study is to measure radiation leakage of Siemens Primus Plus and Siemens Artiste linear accelerators in electron mode and to compare the leakage level with that recommended by the International Electrotechnical Commission (IEC) standard. Materials and Methods: In this assessment, Siemens Primus Plus linear accelerator with 10 cm × 10 cm, 15 cm × 15 cm, and 25 cm × 25 cm applicators was used. The radiation leakage in lateral and vertical directions was measured for Siemens Primus Plus and Siemens Artiste linear accelerators. Results: Data derived from radiation leakage measurement for Siemens Primus Plus and Siemens Artiste linear accelerators in lateral direction from the field edge and in vertical direction from the applicator were reported. The radiation leakage data were then compared with the IEC standard to evaluate in‑air field leakage. Conclusion: Comparing the radiation leakage level from fields with the IEC standard for two applicators, the maximum that was occurred for 12 MeV electron beam and applicator size of 10 cm × 10 cm in Siemens Artiste linear accelerator was 2.3%, which is less than the IEC’s recommended limit of 10%. It is concluded that the leakage amount is much less than the specified limit and that both of the linear accelerators have high level of safety. Considering the measurement stage, it also needs to be noted that the beam angle affected the radiation leakage level from field edge, and in 25° angle, it is higher than in 0° angle. Comparing radiation leakage from the right side of the field for the two linear accelerators, the amount of leakage for Siemens Primus Plus linear accelerator is more than Siemens Artiste linear accelerator.

Keywords

, Applicator, electron beam, linear accelerator, radiation leakage, radiotherapy
برای دانلود از شناسه و رمز عبور پرتال پویا استفاده کنید.

@article{paperid:1078591,
author = {Zohur Karbaf Barbari, Seyede Motahareh and Hadizadeh Yazdi, Mohammad Hadi and مهدی قربانی and سارا عبداللهی},
title = {Assessment of radiation leakage from treatment applicator of Siemens Primus Plus and Siemens Artiste linear accelerators},
journal = {Journal of Cancer Research and Therapeutics},
year = {2018},
volume = {15},
number = {1},
month = {January},
issn = {0973-1482},
keywords = {Applicator; electron beam; linear accelerator; radiation leakage; radiotherapy},
}

[Download]

%0 Journal Article
%T Assessment of radiation leakage from treatment applicator of Siemens Primus Plus and Siemens Artiste linear accelerators
%A Zohur Karbaf Barbari, Seyede Motahareh
%A Hadizadeh Yazdi, Mohammad Hadi
%A مهدی قربانی
%A سارا عبداللهی
%J Journal of Cancer Research and Therapeutics
%@ 0973-1482
%D 2018

[Download]